Grade I: Reverse Osmosis Grade (ASTM Type III )
Grade II: General Lab Testing Grade(Deionized water, ASTM Type II )
Grade IIIA: Analytical Grade (Deionized and free organic water, ASTM Type I )
Grade IIIB: Bioresearch Grade (Deionized, free organic & pyrogent free water, ASTM Type I )
Specification of purify water provided by CIF
| Type of water |
Specification |
Purpose of use |
Grade I: Reverse Osmosis Grade |
With ionization and fine filtration, 85-95% of ions, 99% of microorganism are excluded. |
Washing scientific apparatus |
Grade II: General Lab Testing Grade (Deionized water) |
Deionized water with electrical resistance of 18.3 megaohm-cm* |
General use in laboratories, Stock solution preparation, Culture media preparation, Lab work in histology, serology, parasitology, hematology, microbiology |
Grade IIIA: Analytical Grade (Deionized and free organic water) |
Deionized water with electrical resistance of 18.3 megaohm-cm* and total organic compound (TOC) less than 10 ppb** |
For analytical works which require high accuracy, e.g. HPLC, FPLC, IC, GC,GC-MS, TOC, AAS, ICP/MS, AAA, Bod |
Grade IIIB: Bioresearch Grade (Deionized, free organic, & pyrogent free water) |
Deionized water with electrical resistance of 18.3 megaohm-cm and total organic compound (TOC) less than 10 ppb. Pyrogen (bacterial endotoxins) contamination is less than 0.005EU/ml. |
Cell and tissue culture DNA and IVF research, Monoclonal antibody, other life science application |
*measured at 25°C
**TOC and metal ion can rapidly increase due to improper handle of the purified water, e.g. storage in plastic or glass containers.